The
equipment is PEFC type short stack evaluation test
equipment consisted of gas supply line, moisturizing
line, cooling line, cell performance line and safety
line. |
| ¡@ |
Features
- Standalone and
compact equipment
- Mixed gas with
optional mixing ratio can be produced by setting
total volume and mixing ratio of each gas. Cell is
protected by over-flow protection function.
- Gas utilizing ratio
constant control. Tafel measurement (IV
characteristics), Resistance measurement (option)
- Stable
humidification control, Humidification unit water
temperature (+30¢XC to 90¢XC), Long term stable
operation by humidification unit self-cleaning
system, High-speed humidification response (within 3
min - option)
- High-speed response
cooling system
- Resistance
measurement at conduction of fuel cell (option)
- Various options
including pressure test, humidity measurement, etc.
- Automatic operation
by exclusive control software (CISAS-EX) for
controlling of parameters such as gas flow, bubbler
temperature, pipe temperature control, load current,
etc. for evaluation tests.
¡@ |
Specifications
Gas
supply line
¡@
|
Gas |
Maximum flow |
Control range |
|
Hydrogen (H2) |
30NL to min
|
0.6 to 30NL/min |
|
Carbon dioxide (CO2)
|
15NL/min
|
0.3 to15NL/min |
|
Mixed gas |
1NL/min |
0.02 to 1NL/min |
|
Air |
100NL/min |
3.0 to 100NL/min |
Purge nitrogen
Fuel side
Air side |
About 10NL/min each |
No control |
Flow setting accuracy: ¡Ó1%
of F.S.
Flow control range: 2% of
maximum flow to maximum flow
Repetition accuracy: Within
¡Ó0.25% (indicated value)
MFC response time: Within 1
second
Moisturizing
control line
Moisturizing range: Within
room temperature+30¢XC to 90¢XC¡Ó1¢XC (at stabilized)
Moisturizing capacity: 95%RH
or more
Material: SUS316
Temperature drop at water
supply: Within 1¢XC
Water supply: Automatic
power supply (Ion exchanged water of 6M
or more)
Moisturizing tank automatic
cleaning system: Built-in
Stack
cooling line
Cooling water temperature:
+20¢XC to 90¢XC, ¡Ó1¢XC
Cooling water flow: 1 to
5L/min, ¡Ó1% flow meter range
|
Control system:
|
Inlet
temperature/flow constant control
Inlet temperature/temperature difference
constant control |
Cooling capacity: 1.5KW
Running
pressure control line (option)
Pressure control range:
maximum 500kPa, within ¡Ó5kPa
Pressure response: Changing
by test gas flow/equipment capacity
Operation
control software
Stable software being used
in various industrial applications
Automatic forwarding of
testing items
Excellent performance to
noise
Reduction of CPU load
Exclusive software part for
fuel cell
The software changes process conditions by sequencing
numbers of sheets writing test. After judgment of
stability, it automatically collect data.
(Electronic load unit
control)
(1) Initial activation
automatic operation
(2) Load current sweep test=Tafel
test
Electronic
load (several hundred W to several hundred kW)
As a bias power supply
corresponding to 0V, it can be used in a single cell.
3 ranges of current
settings, high accuracy of ¡Ó0.3% of F.S.
For wide frequency from 5mHz
to 30kHz
For resistance measurement
of IR drop, frequency fixed alternative method,
frequency sweep alternative method
¡@
|
Items |
Specifications |
|
Load mode |
Constant
current/constant resistance/constant power |
|
Discharge voltage
range |
0 to130V |
|
Maximum current |
0 to 1,560A |
|
Maximum load power |
0 to 100kW
|
|
Current setting |
3-range, ¡Ó0.3% of
F.S. |
|
Constant voltage
setting |
0 to 130V/0.01V,
¡Ó0.2% of F.S. |
|
Constant power
setting |
3-range, ¡Ó0.8% of
F.S. |
|
Voltage measurement |
1-range, ¡Ó0.1% of
F.S. |
|
Current measurement |
2-range, ¡Ó0.3% of
F.S. |
|
Resistance
measurement |
3-range,10/100/1000m |
|
|